Procedure Manual for the I-STAT® System

Procedure Manual for the I-STAT® System

Procedure Manual for the i-STAT® System

This Procedure Manual is intended to be a template for the Procedure Manual required by CLIA and laboratory accreditation bodies. This Procedure Manual should be customized for site-specific policies and procedures. This Procedure Manual is not intended to replace the System Manual.

i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions.

ART: 714446-00P REV. DATE: 23-Apr-14 REV. DATE: 23-Apr-14 ART: 714446-00P

PROCEDURE MANUAL FOR THE i-STAT SYSTEM

System Overview ...... 1 Analyzers Analysis Time Cartridges Central Data Station or Data Manager

Supplies and Storage Requirements ...... 1-2 Cartridges Controls

Blood Specimens ...... 2-5 Blood Collection Equipment Blood Volume Suitable Specimens Specimen Labeling Specimen Collection and Handling Criteria for Specimen Rejection Precautions

Procedure for Analysis ...... 5-10 Preparation for Use Procedure for Cartridge Testing Alternative Procedure

Results ...... 11-17 Calculations Displayed Results Suppressed Results Printing and Transmitting Results Reference Ranges and Reportable Ranges Critical Results Interferences

Quality Control ...... 18-23 Daily Procedures Monthly Procedures Periodic Procedures

Calibration ...... 23

Clinical Significance ...... 24-26 Principles of Measurement ...... 27-28

References ...... 29

Logs ...... i-vi

REV. DATE: 23-Apr-14 ART: 714446-00P

SYSTEM OVERVIEW

The i-STAT System incorporates comprehensive components needed to perform blood analysis at the point of care. The System consists of the following primary components: i-STAT 1 Analyzer When a sample-filled i-STAT cartridge is inserted into the i-STAT 1 handheld for analysis, the handheld automatically controls all functions of the testing cycle including fluid movement within the cartridge, calibration and continuous quality monitoring.

Analysis Time q ACT cartridge: to detection of end point - up to 1000 seconds (16.7 min.) q PT/INR cartridge: to detection of end point – up to 300 seconds (5 min.) q cTnI and BNP cartridges: 600 seconds (10 min.) q CK-MB Cartridge: 300 seconds (5 min.) q Other cartridges: typically 130 to 200 seconds Cartridges A single-use disposable cartridge contains microfabricated sensors, a calibrant solution, fluidics system, and a waste chamber. Sensors for analysis of pH, PCO2, PO2, TCO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, creatinine, urea nitrogen (BUN) and hematocrit are available in a variety of panel configurations. Cartridges are also available for Celite-ACT, Kaolin-ACT, PT/INR, Troponin I/cTnI, CK-MB and BNP (Table 1). A whole blood sample of approximately 1 to 3 drops is dispensed into the cartridge sample well, and the sample well is sealed before inserting it into the analyzer.

Central Data Station or Data Manager A dedicated desktop computer with the i-STAT Central Data application provides the primary information management capabilities for the i-STAT System. Downloaders and Downloader/Rechargers for the i-STAT 1 Analyzer allow for transmission of patient records from a widely distributed network of handhelds to the Central Data Station application. Data can be stored, organized, edited, and transferred to a laboratory information system or other computer system. Cartridge usage and efficiency reports can be generated for management of the System.

SUPPLIES and STORAGE REQUIREMENTS Cartridges Cartridges are sealed in individual pouches or portion packs. Store the main supply of cartridges at a temperature between 2 to 8°C (35 to 46°F). Do not allow cartridges to freeze. Cartridges may be stored at room temperature (18 to 30°C or 64 to 86°F) for the time frame indicated on the cartridge box. Cartridges should not be returned to the refrigerator once they have been at room temperature, and should not be exposed to temperatures above 30°C (86°F). If the pouch has been punctured, the cartridge should not be used. Write the date on the cartridge box or individual cartridge pouches to indicate the room temperature expiration date. Cartridges should remain in pouches until time of use. Do not use after the labeled expiration date.

Note: See the Check Temperature Monitor section for information regarding the four-window temperature indicator shipped with cartridges during transit.

1 REV. DATE: 23-Apr-14 ART: 714446-00P

Controls i-STAT Controls for blood gases , electrolytes, and chemistries Store at 2 to 8°C (35° to 46°F). Controls may be stored at room temperature (18 to 30°C or 64 to 86°F) for five days. Do not use after expiration date on the box and ampules. i-STAT Controls for ACT and PT/INR Store at 2 to 8°C (35° to 46°F). Do not use after expiration date on the box and vials. Controls should be used immediately after reconstitution. i-STAT Controls for cTnI , BNP, and CK-MB These controls require no reconstitution or frozen storage. They are stable until the expiration date on the vial label when stored unopened at 2 to 8 °C. Once opened, the i-STAT cTnI, BNP, and CK-MB Controls are stable for 30 days when stored tightly capped at 2 to 8°C.

CLINIQA Liquid Cardiac Marker Control for CK-MB This control requires no reconstitution or frozen storage. It is stable until the expiration date on the vial label when stored unopened at 2 to 8°C. Once opened, CLINIQA Liquid Cardiac Marker Control is stable for 30 days when stored tightly capped at 2 to 8°C.

RNA Medical Hematocrit Control RNA Medical Hematocrit Control is stable until the expiration date stated on the ampule when stored at temperatures of 2 to 25°C. Do not freeze or expose ampules to temperatures greater than 30°C. If stored refrigerated, the control material should be equilibrated to room temperature for at least 4 hours prior to testing.

Electronic Simulator Store at room temperature and protect contact pads from contamination by replacing the plastic cap and placing the Electronic Simulator in its protective case after use.

BLOOD SPECIMENS Blood Collection Equipment Cartridges for Blood Gas /Electrolytes/Chemistries/Hematocrit q Skin puncture: lancet and capillary collection tube (plain, lithium heparin, or balanced heparin for electrolytes and blood gases) q Venipuncture: lithium or sodium heparin collection tubes and disposable transfer device. q Arterial puncture: Plain syringe or blood gas syringe with heparin and labeled for the assays performed or with the least amount of heparin to prevent clotting (10 U heparin/mL of blood) Cartridges for ACT q Skin puncture: not recommended q Venipuncture and arterial puncture: plain plastic syringe without anticoagulant Cartridges for PT/INR q Skin puncture: lancet only needed. Cartridge can be filled directly from the finger. q Venipuncture: plain plastic syringe without anticoagulant. Cartridges for Troponin I/ cTnI and CK-MB q Skin puncture: not recommended. q Venipuncture: lithium or sodium heparin collection tubes and disposable transfer device (e.g. 1 cc syringe and a. Alternately, a plain syringe or plain collection tube and disposable transfer device can be used if the sample is tested within one minute of patient draw. Cartridges for BNP 2 REV. DATE: 23-Apr-14 ART: 714446-00P

q Skin puncture: not recommended. q Venipuncture: plastic EDTA collection tubes and disposable transfer device or plastic EDTA syringe.

Blood Volume See Table 1 below for cartridge volumes.

Table 1: Cartridge Panel Configurations and Blood Volume (Shading denotes calculated values)

MB L - K Gap Cl Na µ

pH iCa Hct Hct Glu Vol. Lact PO2 cTnI BNP Hb ACT BUN Creat BE BE PCO TCO SO2 TCO2 HCO3 PT/INR CK Cartridge

CHEM8+ 95 • • • • • • • • • • • CG8+ 95 • • • • • • • • • • • • • EG7+ 95 • • • • • • • • • • • • EG6+ 95 • • • • • • • • • • • CG4+ 95 • • • • • • • • G3+ 95 • • • • • • • EC8+ 65 • • • • • • • • • • • • • 6+ 65 • • • • • • • EC4+ 65 • • • • • E3+ 65 • • • • G 65 • Crea 65 • ACT 40 • PT/INR 20 • cTnI 17 • CK-MB 17 • BNP 17 •

Suitable Specimens for Cartridges for Blood Gases, Electrolytes, Chemistries, and Hematocrit q Fresh whole blood collected in a plain capillary collection tube or capillary collection tube with balanced heparin. q Fresh whole blood collected in a collection tube with lithium or sodium heparin anticoagulant. Fill collection tubes to capacity. q Fresh whole blood collected in a plain plastic syringe or in a blood gas syringe labeled for the assays to be performed. Fill syringes for correct blood-to-heparin ratio. Suitable Specimens for ACT q Fresh whole blood without anticoagulant collected in a plastic syringe. If from an indwelling line, flush the line with 5mL saline and discard the first 5mL of blood or six dead space volumes of the catheter.

q Fresh whole blood collected in a plastic tube without anticoagulant, clot activators, or serum separators. Device used to transfer sample to cartridge must be plastic. Suitable Specimens for PT/INR q Fresh whole blood without anticoagulant collected in a plastic syringe or plastic evacuated tube without clot activators or serum separators. Device used to transfer sample to cartridge must be plastic.

q Fresh capillary whole blood dispensed directly into the cartridge from the finger. Suitable Specimens for Troponin I/cTnI and CK-MB q Fresh heparinized whole blood or plasma samples collected in syringes or evacuated tubes containing lithium or sodium heparin. Fill collection tube to capacity. q Non heparinized whole blood samples tested within one minute of patient draw collected into a plastic syringe or plastic evacuated tube containing no additives. Suitable Specimens for BNP

3 REV. DATE: 23-Apr-14 ART: 714446-00P

q EDTA whole blood plasma samples collected in plastic syringes or evacuates tubes containing EDTA. Fill collection tubes to capacity.

Specimen Labeling Unless the specimen is analyzed immediately after collection and then discarded, the specimen container must be labeled with the following information: Patient name, sex, age Patient ID number Time and date of collection Phlebotomist ID Doctor’s name

Specimen Collection and Handling In-Dwelling Line Back flush line with sufficient amount of blood to remove intravenous solution, heparin, or medications that may contaminate the sample. Recommendation: five to six times the volume of the catheter, connectors, and needle. If collecting sample for ACT, clear the line first with 5mL saline and discard the first 5mL of blood or six dead space volumes of the catheter. Arterial Specimens For cartridge testing of blood gases, electrolytes, chemistries, and hematocrit, fill a plain syringe or fill a blood gas syringe, labeled for the assays to be performed, to the recommended capacity, or use the least amount of liquid heparin anticoagulant that will prevent clotting. Under-filling syringes containing liquid heparin will decrease results due to dilution and will decrease ionized calcium results due to binding. For ionized calcium, balanced or low volume heparin blood gas syringes should be used. Do not expose sample to air or PCO2 may decrease, pH may increase and PO2 may decrease if the value is above or increase if the value is below the PO2 of room air (approximately 150 mmHg). For cartridge testing of ACT, use only a plain, plastic syringe without anticoagulant. Mix blood and anticoagulant by rolling syringe between palms for at least 5 seconds each in two different directions, then invert the syringe repeatedly for at least 5 seconds. Discard the first two drops of blood. For blood gas testing, avoid or remove immediately any air drawn into syringe to maintain anaerobic conditions. Test samples collected without anticoagulant immediately. Test samples for ACT and lactate immediately. For pH, blood gases, TCO2 and ionized calcium, test within 10 minutes of collection. If not tested immediately, remix the sample and discard the first two drops of blood from a syringe before testing. Note that it may be difficult to property remix a sample in a 1.0 cc syringe. For other cartridge tests, test sample within 30 minutes of collection. Venous Specimens For cartridge testing of electrolytes, chemistries, and hematocrit, collect sample into an evacuated blood collection tube or a syringe containing sodium, lithium, or balanced heparin anticoagulant. For ionized calcium measurements, balanced heparin or 10 U of sodium or lithium heparin/mL of blood is recommended. Fill tubes to capacity; fill syringes for correct heparin-to- blood ratio. Incomplete filling causes higher heparin-to-blood ratio, which will decrease ionized calcium results and may affect other results. The use of partial – draw tubes (evacuated tubes that are adjusted to draw less than the tube volume, e.g. a 5 mL tube with enough vacuum to draw only 3 mL) is not recommended for blood gas or CHEM8+ cartridges because of the potential for decreased PCO2, HCO3 and TCO2 values. For cartridge testing of ACT or PT/INR, use only a plain, plastic syringe or collection tube containing no anticoagulant. Use a plastic capillary tube, pipette, or syringe to transfer sample from a tube to a cartridge. For glucose cartridge testing, EDTA is also an acceptable anticoagulant. EDTA is the only acceptable anticoagulant for BNP Cartridge testing. Mix blood and anticoagulant by inverting a tube gently at least ten times. Roll a syringe vigorously between the palms for at least 5 seconds each in two different directions, then invert the syringe repeatedly for at least 5 seconds, then discard the first two drops of blood. Note that it may be difficult to properly mix a sample in a 1 cc syringe. Test Sample collected without anticoagulant immediately. Test samples for ACT, lactate and PT/INR immediately. Test samples for pH, PCO2, TCO2 and ionized calcium within 10 minutes of sample draw. If not tested immediately, remix the 4 REV. DATE: 23-Apr-14 ART: 714446-00P

sample before testing and discard the first two drops of blood from a syringe before testing. For the glucose test strip and other cartridge tests, test sample within 30 minutes of collection. Finger and Heelstick Specimens For tests other than PT/INR, wipe away the first drop of blood, which contains excess tissue fluid which can increase the potassium result and decrease other test results. Avoid drawing air into the capillary tube. Use balanced heparin or plain capillary tubes for ionized calcium. Test samples immediately to avoid clotting (especially in neonates). Capillary samples are NOT recommended for ACT, Troponin I/cTnI, CK-MB and BNP.

Criteria For Specimen Rejection q Evidence of clotting q Specimens collected in vacuum tubes with anticoagulant other than lithium or sodium heparin (or EDTA for BNP or glucose cartridges). q Specimens for ACT or PT/INR collected in glass syringes or tubes or with anticoagulant of any kind q Syringe for pH, PCO2, PO2 and TCO2 with air bubbles in sample q Incompletely filled vacuum tube for the measurement of ionized calcium, PCO2, HCO3 or TCO2 q Other sample types such as urine, CSF, and pleural fluid Precautions: Avoid the Following Circumstances q Drawing a specimen from an arm with an I.V. q Stasis (tourniquet left on longer than one minute before venipuncture) q Extra muscle activity (fist pumping) q Hemolysis ( alcohol left over puncture site, or a traumatic draw) q Icing before filling cartridge q Time delays before filling cartridge, especially lactate, ACT, and PT/INR q Exposing the sample to air when measuring pH, PCO2, PO2 and TCO2.

PROCEDURE FOR ANALYSIS Preparation for Use An individual cartridge may be used after standing 5 minutes, in its pouch, at room temperature. An entire box should stand at room temperature for one hour before cartridges are used.

5 REV. DATE: 23-Apr-14 ART: 714446-00P

6 REV. DATE: 23-Apr-14 ART: 714446-00P

7 REV. DATE: 23-Apr-14 ART: 714446-00P

8 REV. DATE: 23-Apr-14 ART: 714446-00P

9 REV. DATE: 23-Apr-14 ART: 714446-00P

Procedure for Cartridge Testing

i-STAT Cartridge

A level surface includes running the handheld in the downloader/recharger.

7. Review results

Alternative Procedure Should the i-STAT System become inoperable for any reason, specimens should be collected and submitted to the laboratory in accordance with the Laboratory Procedure Manual.

10 REV. DATE: 23-Apr-14 ART: 714446-00P

RESULTS Calculations The i-STAT handheld contains a microprocessor that performs all calculations required for reporting results.

Displayed Results Results are displayed numerically with their units. Electrolyte, chemistry and hematocrit results are also depicted as bar graphs with reference ranges marked under the graphs.

Suppressed Results There are three conditions under which the i-STAT System will not display results:

1. Results outside the System’s reportable ranges are flagged with a , indicating that the result is below the lower limit or above the upper limit of the reportable range respectively. (See the table of Reportable Ranges.) The flag indicates that the results for this test were dependant on the result of a test flagged as either > or <. Action: Send specimen(s) to the laboratory for analysis, if necessary. 2. Cartridge results which are not reportable based on internal QC rejection criteria are flagged with ***. Action: Analyze the specimen again using a fresh sample and another cartridge. If the specimen integrity is not in question, the results that are not suppressed should be reported in the usual manner. If the result is suppressed again, send specimen(s) to the laboratory for analysis in accordance with the Laboratory Procedure Manual. 3. A Quality Check message will be reported instead of results if the handheld detects a problem with the sample, calibrant solution, sensors, or mechanical or electrical functions of the handheld during the test cycle. Action: Take the action displayed with the message that identifies the problem. Refer to the i-STAT or i-STAT 1 System Manual’s Troubleshooting section or the “Analyzer Coded Messages” Technical Bulletin if necessary.

Printing and Transmitting Results Printing Results from the i-STAT 1 Analyzer to the Martel Portable Printer or to the i-STAT Printer Without Downloader or Downloader/Recharger 1. Turn printer on if green power light is not on. 2. Align IR windows of handheld and printer. 3. Display results. 4. Press the Print key. 5. Do not move handheld or printer until printing is complete. 6. If printer is not powered from a wall outlet, turn printer off.

With Downloader or Downloader/Recharger 1. Place handheld in Downloader or Downloader/Recharger that is wired to the printer. 2. Display results. 3. Press the Print key. 4. Do not move handheld or printer until printing is complete.

Printing more than one result 1. Turn the handheld on. 2. Press the Menu key. 3. Press 2 for Data Review. 4. Press 7 for List. 5. Scroll through the test records using the ← and → keys. 11 REV. DATE: 23-Apr-14 ART: 714446-00P

6. Press the numbered key for the test record(s). (Press the numbered key again to deselect a record.) 7. Align handheld and printer IR window or place in Downloader or Downloader/Recharger attached to printer. Press the Print key. 8. Do not move handheld or printer until printing is complete. 9. If printer is not powered from a wall unit using the AC adapter, turn printer off.

Transmitting Results from the i-STAT 1 Analyzer to the Data Manager 1. Place handheld in a Downloader or Downloader/Recharger. 2. Do not move handheld while the message “Communication in Progress” is displayed.

12 REV. DATE: 23-Apr-14 ART: 714446-00P

Reference Ranges1,2, Reportable Ranges, and Test Unit Conversions Reference range means the range of test values expected from 95% of fasting individuals presumed to be healthy. Reportable range means the range of test values throughout which the measurement system’s results have been shown to be valid. The following table contains the Reference Ranges (for adults) and Reportable Ranges applicable to the i-STAT System.

Cartridges ANALYTE UNIT REFERENCE RANGE REPORTABLE UNIT CONVERSION RANGE (arterial) (venous)

Sodium mmol/L (mEq/L) 138 – 146 138 - 146 100 – 180 mmol/L x 1 = mEq/L Example: 140 mmol/L = 140 mEq/L

Potassium mmol/L (mEq/L) 3.5– 4.9 3.5 - 4.9 2.0 – 9.0 mmol/L x 1 = mEq/L

Chloride mmol/L (mEq/L) 98 – 109 98 – 109 65 – 140 mmol/L x 1 = mEq/L

BUN mg/dL 8 – 26 8 – 26 3 – 140 mg/dL BUN x 0.357 = mmol urea/L Example: 20 mg/dL BUN = 7.1 mmol urea/L UREA mmol/L 2.9 – 9.4 2.9 – 9.4 1 – 50

Glucose mg/dL 70 – 105 70 – 105 20 – 700 mg/dL x 0.055 = mmol/L Example: g/L 0.70 – 1.05 0.70 – 1.05 0.20 – 7.00 100 mg/dL = 5.55 mmol/L mmol/L 3.9 – 5.8 3.9 – 5.8 1.1 – 38.9 g/L x 5.556 = mmol/L Creatinine mg/dL 0.6 - 1.3 0.6 - 1.3 0.2 - 20.0 mg/dL x 88.4 = µmol/L

µmol/L 53 - 115 53 - 115 18 - 1768

Ionized Calcium mmol/L 1.12 – 1.32 1.12 – 1.32 0.25 – 2.50 mmol/L x 4 = mg/dL mg/dL 4.5 – 5.3 4.5 – 5.3 1.0 – 10.0 Example: 1.13 mmol/L x 4 = 4.52 mg/dL

pH 7.35 – 7.45 7.31 – 7.41 6.50 – 8.20 N/A

PCO2 mmHg 35 – 45 41 – 51 5 – 130 mmHg x 0.133 = kPa Example: kPa 4.67 – 6.00 5.47 – 6.80 0.67 – 17.33 35 mmHg x 0.133 = 4.66 kPa

PO2 mmHg 80 – 105 5 – 800 mmHg x 0.133 = kPa Example: kPa 10.7 – 14.0 0.7 – 106.6 83 mmHg x 0.133 = 11.04 kPa

TCO2 (on the CHEM8+ mmol/L (mEq/L) 23-27 24-29 5-50 mmol/L x 1 = mEq/L cartridge only

13 REV. DATE: 23-Apr-14 ART: 714446-00P

ANALYTE UNIT REFERENCE RANGE REPORTABLE UNIT CONVERSION RANGE (arterial) (venous)

Hematocrit %PCV 38 – 51 38 – 51 10 – 75 % PCV x 0.01 = Volume fraction Fraction 0.38 – 0.51 0.38 – 0.51 0.10 – 0.75 Example: 40% PCV = 0.40 PCV

Lactate mmol/L 0.36 –1.25 0.90-1.70 0.30 – 20.00 mmol/L x 9.01 = mg/dL mg/dL 3.2 – 11.3 8.1–15.3 2.7 – 180.2

HCO3* mmol/L (mEq/L) 22 – 26 23 - 28 1.0 – 85.0 mmol/L x 1 = mEq/L

TCO2* mmol/L (mEq/L) 23 – 27 24 - 29 5 – 50 mmol/L x 1= mEq/L (on all cartridges but CHEM8+)

BE* mmol/L (mEq/L) (-2) – (+3) (-2) – (+3) (-30) – (+30)

Anion Gap* mmol/L (mEq/L) 10 – 20 10 – 20 (-10) – (+99)

sO2* % 95 – 98 0 - 100 % x 0.01 = fraction saturated

Hb* g/dL 12 – 17 12 – 17 3.4 – 25.5 g/dL x 10 = g/L g/L 120 – 170 120 – 170 34 – 255 mmol/L 7 – 11 7 – 11 2.1 – 15.8 Celite ACT seconds 74 – 125 (PREWRM) 74 – 125 (PREWRM) 50 - 1000 84 – 139 (NONWRM) 84 – 139 (NONWRM) Kaolin ACT seconds 74 – 137 (PREWRM) 74 – 137 (PREWRM) 50 - 1000 82 – 152 (NONWRM) 82 – 152 (NONWRM) Prothrombin INR 0.9-8.0# Time/PT Troponin I/cTnI ng/mL (µg/L) 0.00 – 0.03** 0.00 –50.00## ng/mL x 1 = µg/L 0.00 – 0.08*** Creatine ng/mL (µg/L) 0.0 – 3-5**** 0.0-150.0 ng/mL x 1 = µg/L Kinase MB/ CK-MB B-Type pg/mL (ng/L) <15-50**** 15-5000 pg/mL x 1 = ng/L Natriuretic Peptide/BNP

*Calculated values. #Performance characteristics have not been established for INRs above 6.0. **Represents the 0 – 97.5% range of results. Each facility should establish it’s own reference range using the i-STAT cTnI assay. ##Performance characteristics have not been established for cTnI values above 35.00 ng/mL. ***Represents the 0-99% range of results. Each facility should establish it’s own reference range using the i-STAT cTnI assay. ****Represents the 0-95% range of results. Each facility should establish it’s own reference range using the i-STAT assay

14 REV. DATE: 23-Apr-14 ART: 714446-00P

Critical Results3 Critical results are test results that fall outside high and low critical limits that define the boundaries of life-threatening values for a test. Critical results represent an emergency condition and must be reported immediately to the patient’s attending physician or nurse.

ANALYTE (units) ADULT CHILDREN NEONATES low high low high low high Sodium (mmol/L) 120 158 121 156 121 156

Potassium (mmol/L) 2.8 6.2 2.8 6.4 2.8 6.5

Chloride (mmol/L) 75 126 77 121 77 121

TCO2 (mmol/L) 11 40 11 39 – –

Ionized Calcium (mmol/L) 0.78 1.58 0.74 1.57 – –

pH 7.21 7.59 7.21 7.59 – –

PCO2 (mmHg) 19 67 21 66 – –

PO2 (mmHg) 43 – 45 124 37 92

BUN (mg/dL) – 104 – 55 – 55

Glucose (mg/dL) 46 484 46 445 32 328

Creatinine – 7.4 – 3.8 – –

Hematocrit (%PCV) 18 61 20 62 33 71

Troponin I/cTnI

Creatine Kinase MB/ CK-MB

B-Type Natriuretic Peptide/ BNP

15 REV. DATE: 23-Apr-14 ART: 714446-00P

Interferences An interferent is a substance which, if present at significant levels in the blood specimen being analyzed, will produce an error in the result of the analyte being measured.

ANALYTE INTERFERENT INTERFERENT CONCENTRATION EFFECT ON ANALYTE RESULT

Sodium Bromide 37.5 mmol/L Increase (↑) Na

Potassium Bromide 37.5 mmol/L Increased rate of star (***) outs.

Chloride Acetylcysteine 10.2 mmol/L Decrease (↓) Cl

Bromide 37.5 mmol/L Increase (↑) Cl

Bromide (therapeutic) 2.5 mmol/L Increase (↑) Cl

Salicylate 4.34 mmol/L Increase (↑) Cl

Thiocyanate 6.9 mmol/L Increase (↑) Cl

Ionized Calcium Acetominophen 1.32 mmol/L Decrease (↓) iCa

Magnesium 1.0 mmol/L Increase (↑) iCa by 0.04 mmol/L

Acetylcysteine 10.2 mmol/L Decrease (↓) iCa

Bromide 37.5 mmol/L Increase (↑) iCa

Lactate 6.6 mmol/L Decrease (↓) iCa by 0.07 mmol/L

Salicylate (therapeutic) 0.5 mmol/L Decrease (↓) iCa by approx. 0.03 mmol/L

Salicylate 4.34 mmol/L Decrease (↓) iCa Thiocyanate 6.9 mmol/L Decrease (↓) iCa. Use Another Method.

Glucose Acetominophen 1.32 mmol/L Increase (↑) glucose

Acetylcysteine 10.2 mmol/L Decrease (↓) glucose

Bromide 37.5 mmol/L Decrease (↓) glucose

Bromide (therapeutic) 2.5 mmol/L Decrease (↓) glucose

pH pH: per 0.1 pH units below 7.4 @ 37°C Decrease (↓) glucose by 0.9 mg/dL (0.05 mmol/L)

pH: per 0.1 pH units above 7.4 @ 37°C Increase (↑) glucose by 0.8 mg/dL (0.04 mmol/L)

Oxygen PO2 less than 20 mmHg @ 37°C May decrease (↓) glucose

Hydroxyurea 0.92 mmol/L Increase (↑) glucose

Thiocyanate 6.9 mmol/L Decrease (↓) glucose by approx. 7 mg/dL

Lactate Bromide 37.5 mmol/L Decrease (↓) lactate

Hydroxyurea 0.92 mmol/L Increase (↑) lactate. Use Another Method. Glycolic Acid 10.0 mmol/L Increase (↑) lactate. Use Another Method.

16 REV. DATE: 23-Apr-14 ART: 714446-00P

BUN/Urea Bromide 37.5 mmol/L Increased rate of star (***) outs Hydroxyurea 0.92 mmol/L Increase (↑) BUN/Urea results

Creatinine Acetaminophen 1.32 mmol/L Increase (↑) creatinine

Ascorbate 0.34 mmol/L Increase (↑) creatinine by 0.3 mg/dL

Bromide (therapeutic) 2.5 mmol/L Increase (↑) creatinine

Increase (↑) creatinine by 6.9% per 10 <2 mg/dL PCO2 Above 40 mmHg mmHg PCO2

Decrease (↓) creatinine by 6.9% per 10 Below 40 mmHg mmHg PCO2

Decrease (↓) creatinine by 3.7% per 10 >2 mg/dL PCO2 Above 40 mmHg mmHg PCO2

Below 40 mmHg Increase (↑) creatinine by 3.7% per 10 mmHg PCO2

Hydroxyurea 0.92 mmol/L Increase (↑) Creatinine. Use Another Method.

Acetylcysteine 10.2 mmol/L Increase (↑) creatinine

Creatine 0.382 mmol/L Increase (↑) creatinine by 0.2 mg/dL

Glycolic Acid 10.0 mmol/L Decrease (↓) creatinine. Use Another Method.

Hematocrit White Blood Count Greater than 50,000 WBC/µL May Increase (↑) hematocrit (WBC) For measured Hct<40% Total Protein For each g/dL below 6.5 Decrease (↓) Hct by 1% PCV For each g/dL above 8.0 Increase (↑) Hct by 1% PCV

For measured Hct≥40% For each g/dL below 6.5 Decrease (↓) Hct by 0.75% PCV For each g/dL above 8.0 Increase (↑) Hct by 0.75% PCV

Lipids Abnormally high Increase (↑) Hct Bromide 37.5 mmol/L Increased rate of star (***) outs

Celite ACT Aprotinin Falsely extends Celite ACT times

® PCO2 Propofol (Diprovan ) For patients administered propofol or thiopental sodium, i-STAT recommends the Thiopental Sodium use of G3+, CG4+, CG8+, EG6+, and EG7+ cartridges, which are free from clinically significant interference at all relevant therapeutic doses. i-STAT does not recommend the use of EC8+ cartridges for patients receiving propofol or thiopental sodium.

PT/INR Cubicin® (daptomycin Falsely extends prothrombin time (PT) for injection) and INR

Chlorhexidine May falsely extend prothrombin time (PT) Gluconate and INR

17 REV. DATE: 23-Apr-14 ART: 714446-00P

QUALITY CONTROL Daily Procedures Handheld Verification Verify the performance of each handheld in the i-STAT System using the internal or external Electronic Simulator every 24 hours of use, or as needed for regulatory compliance. In the USA, verification is required every 8 hours for blood gases, hematocrit, ACT, and PT/INR. Action: If PASS is displayed on the handheld screen (after using the external Electronic Simulator): q Remove the external Electronic Simulator after the LCK or Simulator Locked message disappears from the display screen. q Transmit the result to the Central Data Station. q Use the handheld as required. Note: If the internal Electronic Simulator is used, the “PASS” message will not be displayed on the handheld screen. The “PASS” record will appear in the handheld’s stored results for transmission to the Central Data Station.

Remedial Action: If FAIL is displayed on the analyzer screen: q Repeat the procedure with the same external Electronic Simulator or rerun the cartridge if the internal Electronic Simulator is being used. If PASS is displayed use the handheld as required. q If FAIL is displayed repeat the procedure with a different external Electronic Simulator.

If PASS is displayed with the second external Electronic Simulator: q Use the handheld as required. q Deliver the questionable external Electronic Simulator to the i-STAT System Coordinator.

If FAIL is displayed with the second external Electronic Simulator: q DO NOT analyze patient samples with the handheld. q Transmit the results to the Central Data Station. q Deliver the faulty handheld to the i-STAT System Coordinator. q Record the failure in the i-STAT QC Log along with the action taken.

Verification of Cartridge Storage Conditions Refrigerated Cartridges

q Verify that the cartridges stored in the refrigerator are all within the expiration date printed on the boxes. Deliver any expired cartridges to the i-STAT System Coordinator. q Verify that the refrigerator did not exceed the limits of 2 to 8°C (35 to 46°F). q Document in the i-STAT QC Log. Action: If the temperature of the cartridge storage refrigerator is within the range of 2 to 8°C (35 to 46°F) use cartridges as required. Remedial Action: If the temperature is outside the range of 2 to 8 °C (35 to 46 °F), notify the i-STAT System Coordinator immediately. Record the QC failure in the i-STAT QC Log along with the actions taken as described in Section 14 (Quality Control) of the i-STAT 1 System Manual.

18 REV. DATE: 23-Apr-14 ART: 714446-00P

Room Temperature Cartridges

q Verify that all boxes of cartridges at room temperature have been out of the refrigerator less than the time frame indicated on the cartridge box. Deliver any expired cartridges to the i-STAT System Coordinator. q Verify that room temperature has not exceeded 30°C. q Document in the i-STAT QC log. Action: If the measured temperature of the room has been continuously below 30°C (86°F) use cartridges as required. Remedial Action: If the measured room temperature has exceeded 30°C (86°F) for any period of time: q Quarantine the cartridges. q Notify the i-STAT System Coordinator immediately. q DO NOT USE the cartridges. q Record the out-of-control event in the i-STAT QC Log and the action taken.

Monthly Procedures Print Electronic Simulator Results Print a copy of the Electronic Simulator results from the Central Data Station. Include the report in the i-STAT QC Log.

q CDS version 5 and above: Click on the Simulator Viewer.

Print Control Fluid Analysis Results Print results for any control fluids analyzed from the Central Data Station. Include the report in the i-STAT QC Log.

q CDS version 5 and above: Click on the Control Results Viewer.

Periodic Procedures for Cartridges For acceptance of newly received cartridge lots, check the Temperature Monitor and perform integrity testing.

Check Temperature Monitor i-STAT cartridges are shipped refrigerated with a four-window indicator to monitor temperature during transit. Note: All control and calibration verification materials, except for those shipped on dry ice, will also include a four-window indicator to monitor temperature during transit.

Action: q Fill out the record of receipt and forward materials to refrigerator. q If all windows are white or if only the A or B windows are blue or the 1 or 2 windows are red, then transit temperatures were satisfactory and the cartridges can be used. Remedial Action: If the C or D windows are blue, or the 3 or 4 windows are red: q Quarantine the suspect cartons. q Notify the i-STAT System Coordinator immediately. q DO NOT USE cartridges from the suspect cartons. q Record the out-of-control event in the i-STAT QC Log.

19 REV. DATE: 23-Apr-14 ART: 714446-00P

Integrity Testing From each lot of blood gas/chemistry cartridges received, use a representational number of cartridges to analyze i-STAT or TriControls Level 1 and 3 Controls. For CHEM8+ cartridges, analyze i-STAT CHEM8+ or TriControls Control Levels 1 and 3. If not using TriControls and hematocrit is included on the cartridge type being tested, also analyze RNA Medical Hematocrit Control Levels 1 and 3. For ACT cartridges, analyze i-STAT Level 1 and Level 2 ACT Controls. For PT/INR cartridges, analyze i-STAT Level 1 and Level 2 PT Controls. For cTnI cartridges, analyze i-STAT Level 1 and Level 3 cTnI Controls. For CK-MB cartridges, analyze i-STAT Level 1 and Level 3 CK-MB Controls or CLINIQA Cardiac Marker Level 1 and 3 controls. For BNP cartridges, analyze i-STAT BNP Control Levels 1 and 3. Use any verified analyzer for control testing. If available, transmit the results to the Data Manager. Use the expected values published in the Value Assignment Sheet to verify the integrity of the cartridges.

Note: the above information is not a manufacturer’s system instruction; it is a suggestion to comply with regulatory requirements.

Procedure for testing cartridges with i-STAT Level 1 and Level 3 Controls: 1. Prior to testing cartridges that measure PO2, ampules should stand at room temperature a minimum of 4 hours before use. When testing other cartridges (G, Crea, E3+, EC4+, 6+, or EC8+), ampules may be used once the fluid has reached room temperature, approximately 30 minutes for individual ampules. For best results, ampules, cartridges, and handhelds should be at the same temperature. When using cartridges that contain sensors for measuring ionized calcium, pH, PCO2, or PO2 (G3+, EG6+, EG7+, CG4+, CG8+, or EC8+), a separate ampule must be used for each cartridge being tested; if these sensors are not present (i.e., the 6+ cartridge), the contents of one ampule may be used to fill more than one cartridge as long as the cartridges are filled and inserted into a handheld within 10 minutes of opening the ampule. 2. Immediately before use, shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To shake, hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the temperature of the solution. If necessary, tap the tip of the ampule to send solution back into the bottom section of the ampule. Protect fingers with gauze, tissue, or glove, or use an ampule breaker to snap off the tip of the ampule at the neck. 3. Immediately transfer the solution from the ampule into a plain capillary tube or plain syringe, and then immediately transfer the solution into a cartridge. Immediately seal the cartridge and insert it into a handheld. It is important not to expose the solution to room air since this will alter the results. q When using a capillary tube, fill from the bottom of the ampule. Avoid drawing solution from the surface by covering the far end of the tube as it is inserted into the ampule. Once the open end of the tube rests at the bottom of the ampule, uncover the other end to allow filling by capillary action. q When using a syringe (1cc or 3cc syringes with 16 to 20 gauge needles are recommended), slowly draw approximately 1 mL of solution from the bottom of the ampule. If air is trapped between the leading edge of the solution and the plunger, do not invert the syringe to expel it; this will not affect solution near the tip of the syringe. If air bubbles are continually drawn into the syringe, or if a bubble is trapped near the tip of the syringe, discard the ampule and syringe and use a fresh ampule and syringe. Expel one or two drops from the syringe before filling the cartridge. q Do not use solution left in the syringe, ampule, or capillary tube for additional testing of the cartridges that contain sensors for ionized calcium, pH, PCO2, or PO2. However, cartridges without these sensors may be tested with remaining fluids if within 10 minutes of opening the ampule. 4. Compare results to the value assignment sheet ranges. Check that the lot number on the control ampule matches the lot number on the package insert and that the software version listed on the insert matches the software installed in the handheld. If all results are within expected ranges, use the cartridges as needed. If available, transmit the results to the Data Manager.

Procedure for testing cartridges with TriControls Level 1 and Level 3 Controls: 1 Prior to testing cartridges that measure PO2, ampules should stand at room temperature a minimum of 4 hours before use. When testing other cartridges (G, Crea, E3+, EC4+, 6+, EC8+, or CHEM8+), ampules may be used once the fluid has reached room temperature, approximately 30 minutes for individual ampules. For best results, ampules, cartridges, and handhelds should be at the same temperature. When using cartridges that contain sensors for measuring ionized calcium, pH, PCO2, or PO2 (G3+, EG6+, EG7+, CG4+, CG8+, EC8+, or CHEM8+), a separate ampule must be used for each cartridge being tested; if these sensors are not present (i.e., the 6+ cartridge), the contents of one ampule may be used to fill more than one cartridge as long as the cartridges are filled and inserted into a handheld within 10 minutes of opening the ampule. 20 REV. DATE: 23-Apr-14 ART: 714446-00P

2 Immediately before use, shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To shake, hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the temperature of the solution. If necessary, tap the tip of the ampule to send solution back into the bottom section of the ampule. Protect fingers with gauze, tissue, or glove, or use an ampule breaker to snap off the tip of the ampule at the neck. 3 Immediately transfer the solution from the ampule into a plain capillary tube or plain syringe, and then immediately transfer the solution into a cartridge. Immediately seal the cartridge and insert it into a handheld. It is important not to expose the solution to room air since this will alter the results. q When using a capillary tube, fill from the bottom of the ampule. Avoid drawing solution from the surface by placing a finger over the far end of the tube as it is inserted into the ampule. Once the open end of the tube rests at the bottom of the ampule, uncover the other end to allow filling by capillary action. q When using a syringe (1cc or 3cc syringes with 16 to 20 gauge needles are recommended), slowly draw approximately 1 mL of solution from the bottom of the ampule. If air is trapped between the leading edge of the solution and the plunger, do not invert the syringe to expel it; this will not affect solution near the tip of the syringe. If air bubbles are continually drawn into the syringe, or if a bubble is trapped near the tip of the syringe, discard the ampule and syringe and use a fresh ampule and syringe. Expel one or two drops from the syringe before filling the cartridge. q Do not use solution left in the syringe, ampule, or capillary tube for additional testing of the cartridges that contain sensors for ionized calcium, pH, PCO2, or PO2. However, cartridges without these sensors may be tested with remaining fluids if within 10 minutes of opening the ampule. 4. Compare results to the Value Assignment Sheet ranges. Ensure that the lot number printed on the Value Assignment Sheet matches the lot number on the label of the ampule and that the software version above the target value table matches the software version in the handheld. If all results are within expected ranges, use the cartridges as needed. If available, transmit the results to the Data Manager.

Procedure for testing CHEM8+ cartridges with i-STAT CHEM8+ controls. 1. Equilibrate the ampule for approximately 30 minutes at room (ambient) temperature. 2. Immediately before use, shake the ampule vigorously for 5-10 seconds to equilibrate the liquid and gas phases. To shake, hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the temperature of the solution. If necessary, tap the tip of the ampule to send solution back into the bottom section of the ampule. 3. Protect fingers with gauze, tissue or glove, or use an ampule breaker to snap off the tip of the ampule at the neck. 4. Immeditely transfer the solution from the ampule into a plain capillary tube or plain syringe, and then immediately transfer the solution into a cartridge. 5. Immediately seal the cartridge and insert it into a handheld – it is important not to expose the solution to room air since this will alter the results. q When using a capillary tube, fill from the bottom of the ampule. Avoid drawing solution from the surface by covering the far end of the tube as it is inserted into the ampule. Once the open end of the tubes at the bottom of the ampule, uncover the other end to allow filling by capillary action. q When using a syringe (1 cc or 3 cc syringes with 16-20 gauge needles are recommended), slowly draw approximately 1 mL of solution from the bottom of the ampule. If air is trapped between the leading edge of the solution and the plunger, do NOT invert the syringe to expel it; this will not affect solution near the tip of the syringe. If air bubbles are continually drawn into the syringe, or if a bubble is trapped near the tip of the syringe, discard the ampule and syringe and use a fresh ampule and syringe. Expel one or two drops from the syringe before filling the cartridge. q Do not use solution left in the syringe, ampule, or capillary tube for additional testing of CHEM8+ cartridges. 6. Compare results to the value assignment sheet ranges. Check that the lot number on the control ampule matches the lot number on the package insert that the software version listed on the insert matches the software installed in the handheld. If all results are within expected ranges, use the cartridges as needed. If available, transmit the results to the Data Manager.

21 REV. DATE: 23-Apr-14 ART: 714446-00P

Procedure for testing cartridges with RNA Medical Hematocrit controls 1. If stored refrigerated, the control material should be equilibrated to room temperature for at least four (4) hours prior to use. If stored at room temperature, no equilibration of the control material is necessary. 2. Gently invert the ampule to mix the solution. Tap the ampule to restore the liquid to the bottom of the ampule. 3. Open the ampule by snapping off the tip at the neck. Use gauze, tissue, gloves, or an appropriate ampule opener to protect fingers from cuts. 4. Fill and seal a cartridge and insert immediately into the handheld. Note: the Control option from the Quality Tests menu must be used on the i-STAT 1 Analyzer for RNA Medical Hematocrit Control. 5. Compare the i-STAT System Hematocrit result to the value assignment sheet ranges. Note: the only value assigned to this fluid is for hematocrit. All other cartridge analyte results obtained with this control material should be ignored. 6. If available, transmit results to the Data Manager.

Remedial Action: If any results are outside the published expected ranges: q DO NOT USE cartridges from the suspect lot. q Quarantine the suspect lot. q Notify the i-STAT System Coordinator immediately. q Record the QC failure in the i-STAT QC Action Log along with the action taken.

Procedure for testing cartridges with i-STAT Level 1 and Level 2 ACT or PT/INR Controls 1. Prior to use, allow one vial each of the lyophilized plasma and calcium chloride reconstituting fluid to stand at room temperature for a minimum of 45 minutes. 2. Remove the cap and stopper from the vials and pour the entire contents of the calcium chloride vial into the lyophilized plasma vial. Place the stopper back on the reconstituted vial. 3. Allow the vial to sit for 1 minute and then mix the contents by swirling gently for 1 minute, then inverting slowly for 30 seconds. 4. Use a plastic pipette, syringe, or capillary tube without anticoagulant to transfer the solution to an ACT cartridge. 5. Immediately seal the cartridge and insert it into a handheld. This process must be completed within 30 seconds of the complete reconstitution of the control sample. 6. Compare results to the value assignment sheet ranges. If results are within the expected ranges, use the cartridges as needed. If available, transmit results to the Data Manager.

Remedial Action: If any results are outside the published expected ranges:

q DO NOT USE cartridges from the suspect lot. q Quarantine the suspect lot. q Notify the i-STAT System Coordinator immediately. q Record the QC failure in the i-STAT QC Action Log along with the action taken.

Procedures for for testing cartridges with i-STAT cTnI, BNP, and CK-MB controls 1. i-STAT cTnI, BNP, and CK-MB Controls are ready-to-use liquid control requiring no reconstitution or frozen storage. They are stable until the expiration date on the vial label when stored unopened at 2-8°C. Once opened, the i-STAT cTnI, BNP, and CK-MB Controls are stable for 30 days when stored tightly capped at 2-8°C. 2. Access the Control option under Quality Tests in the Administration Menu. Enter the required information. The handheld allows 15 minutes (or the customized timeout period) to insert the cartridge after the last data entry. 3. Immediately before use, gently mix the contents of the control vial to ensure homogeneity. Avoid foaming of the sample. 4. Open the vial and transfer a drop of the solution into the i-STAT cartridge using a plain capillary tube, plain syringe, or plastic transfer pipette. Tightly recap the control vial and store it at 2-8°C. 22 REV. DATE: 23-Apr-14 ART: 714446-00P

5. Seal the cartridge and immediately insert it into the i-STAT 1 Analyzer. 6. Compare the result to the Value Assignment Sheet value. Always ensure that the lot number and software revision on the Value Assignment Sheet matches the lot number of the vial in use and the software revision in the handheld. Should results fall outside the range, refer to the Troubleshooting section of i-STAT 1 System Manual section 14 (Quality Control). 7. If available, transmit results to the Data Manager. Remedial Action If any results are outside the published expected ranges: q DO NOT USE cartridges from the suspect lot. q Quarantine the suspect lot. q Notify the i-STAT System Coordinator immediately. q Record the QC failure in the i-STAT QC Action Log along with the action taken.

Procedures for for testing CK-MB cartridges with CLINIQA Cardiac Marker controls for i-STAT 1. CLINIQA Liquid QC Cardiac Marker Control is a ready-to-use liquid control requiring no reconstitution or frozen storage. It is stable until the expiration date on the vial label when stored unopened at 2-8°C. Once opened, CLINIQA Liquid QC Cardiac Marker Control is stable for 30 days when stored tightly capped at 2-8°C. 2. Immediately before use, gently mix the contents of the control vial to ensure homogeneity. Avoid foaming of the sample. 3. Open the vial and transfer a drop of the solution into the i-STAT CK-MB cartridge using a plain capillary tube, plain syringe, or plastic transfer pipette. Tightly recap the control vial and store it at 2-8°C. 4. Seal the cartridge and immediately insert it into the i-STAT 1 Analyzer. 5. Compare the i-STAT CK-MB result to the Value Assignment Sheet value. Always ensure that the lot number and software revision on the Value Assignment Sheet matches the lot number of the vial in use and the software revision in the handheld. Should results fall outside the range, refer to the Troubleshooting section of i-STAT 1 System Manual section 14 (Quality Control). 6. If available, transmit results to the Data Manager. Remedial Action If any results are outside the published expected ranges: q DO NOT USE cartridges from the suspect lot. q Quarantine the suspect lot. q Notify the i-STAT System Coordinator immediately. q Record the QC failure in the i-STAT QC Action Log along with the action taken.

CALIBRATION For cartridges, calibration is automatically performed as part of the test cycle on each cartridge type, except coagulation and immunoassay cartridges. Operator intervention is not necessary.

CLINICAL SIGNIFICANCE

23 REV. DATE: 23-Apr-14 ART: 714446-00P

Analyte Some Causes of Some Causes of Increased Values Decreased Values

Sodium Dehydration Dilutional hyponatremia Diabetes insipidus (cirrhosis) Salt poisoning Depletional hyponatremia Skin losses Syndrome of inappropriate ADH Hyperaldosteronism CNS disorders

Potassium Renal glomerular disease Renal tubular disease Adrenocortical insufficiency Hyperaldosteronism Diabetic Ketoacidosis (DKA) Treatment of DKA Sepsis Hyperinsulinism In vitro hemolysis Metabolic alkalosis Diuretic therapy

Chloride Prolonged diarrhea Prolonged vomiting Renal tubular disease Burns Hyperparathyroidism Salt-losing renal disease Dehydration Overhydration Thiazide therapy

Ionized Dehydration Hypoparathyroidism Calcium Hyperparathyroidism Early neonatal hypocalcemia Malignancies Chronic renal disease Immobilization Pancreatitis Thiazide diuretics Massive blood transfusions Vitamin D intoxication Severe malnutrition

BUN Impaired renal function Pregnancy Prerenal azotemia (e.g. shock) Severe liver insufficiency Postrenal azotemia Overhydration GI bleeding Malnutrition High protein diet

Glucose Diabetes mellitus Insulinoma Pancreatitis Adrenocortical insufficiency Endocrine disorders (e.g. Cushing’s syndrome) Hypopituitarism/Massive liver Drugs (e.g. steroids, thyrotoxicosis) disease Chronic renal failure Ethanol ingestion/Reactive Stress hypoglycemia IV glucose infusion Glycogen storage disease

Creatinine Impaired renal function

24 REV. DATE: 23-Apr-14 ART: 714446-00P

Lactate Hypoxia (shock, hypovolumia, left ventricular failure); diabetes mellitus, neoplasia, liver disease; drug or toxins (ethanol, methanol, salicylates); glycolic acid as a product of ethylene glycol metabolism

pH Respiratory alkalosis Respiratory acidosis Metabolic alkalosis Metabolic acidosis

PCO2 Acute Respiratory Acidosis: Respiratory alkalosis: • Depression of respiratory center • Increased stimulation of respirator center • Suppressed neuromuscular system Hypermetabolic states • Pulmonary disorders • Mechanical hyperventilation • Inadequate mechanical ventilation • Chronic respiratory acidosis Compensation in metabolic acidosis • Decreased alveolar ventilation • Hypoventilation Compensation in metabolic alkalosis

PO2 Breathing oxygen-enriched air Carbon-monoxide exposure Pulmonary disorders Myocardial infarction Congestive heart failure

HCO3 and Primary metabolic alkalosis Primary metabolic acidosis TCO2 Primary respiratory acidosis Primary respiratory alkalosis

Hematocrit Dehydration Hemolytic anemias Burns Iron deficiency Impaired ventilation Marrow depression Renal disorders Blood loss

ACT Celite Administration of heparin for medical or surgical procedures. Administration of aprotinin.

PT/INR Administration of oral anticoagulant therapy.

25 REV. DATE: 23-Apr-14 ART: 714446-00P

ACT Kaolin Administration of heparin for medical or surgical procedures.

cTnI Myocardial Infarction Rare antibodies to troponin or its Coronary vasospasm circulating complexes Cardiac contusion/trauma Rhythm disturbance (SVT, AF) Chemotherapy (ex. Adriamycin) Myocarditis/pericarditis Infiltrative diseases (ex. Amyloidosis, sarcoidosis, hemochromatosis, connective tissue disease) Congestive heart failure Heart transplantation Cardiac procedures (PTCA, DC cardioversion) Intracranial hemorrhage/stroke Pulmonary embolism Pulmonary hypertension Chronic renal insufficiency Sepsis Strenuous exercise Certain drug ingestions

CK-MB Myocardial Infarction Lean muscle mass Coronary vasospasm Cardiac contusion/trauma Myocarditis/pericarditis Infiltrative diseases (ex. Amyloidosis, sarcoidosis, hemochromatosis, connective tissue disease) Cardiac procedures (PTCA, DC cardioversion) Intracranial hemorrhage/stroke Pulmonary embolism Pulmonary hypertension Chronic renal insufficiency Sepsis Strenuous exercise Certain drug ingestions ( cocaine ) Skeletal muscle disease

BNP Congestive heart failure Obesity (BMI>30 Kg/m2) Chronic obstructive pulmonary disease (COPD) Flash pulmonary edema Asthma (elevation may be delayed) Pulmonary hypertension Cor pulmonale Pulmonary embolism Acute coronary syndrome Chronic renal failure Age Female sex

26 REV. DATE: 23-Apr-14 ART: 714446-00P

PRINCIPLES OF MEASUREMENT

Sodium, Potassium, Chloride, Ionized Calcium, pH, and PCO2 are measured by ion-selective electrode potentiometry. Concentrations are calculated from the measured potential through the Nernst equation.

Urea is first hydrolyzed to ammonium ions in a reaction catalyzed by the enzyme urease. The ammonium ions are measured by an ion-selective electrode and the concentration is calculated from the measured potential through the Nernst equation.

Glucose is measured amperometrically. Oxidation of glucose, catalyzed by the enzyme glucose oxidase, produces hydrogen peroxide. The liberated hydrogen peroxide is oxidized at an electrode to produce an electric current which is proportional to the glucose concentration.

Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed by the enzyme creatine amidinohydrolase. The oxidation of sarcosine, catalyzed by the enzyme sarcosine oxidase , produces hydrogen peroxide. The liberated hydrogen peroxide is oxidized at the platinum electrode to produce a current which is proportional to the creatinine concentration.

Lactate is measured amperometrically. The enzyme lactate oxidase, immobilized in the lactate biosensor, selectively converts lactate to pyruvate and hydrogen peroxide. The liberated hydrogen peroxide is oxidized at the platinum electrode to produce a current which is proportional to the lactate concentration.

PO2 is measured amperometrically. The oxygen sensor is similar to a conventional Clark electrode. Oxygen permeates through a gas permeable membrane from the blood sample into an internal electrolyte solution where it is reduced at the cathode. The oxygen reduction current is proportional to the dissolved oxygen concentration.

Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.

ACT is determined amperometrically. The conversion of a thrombin substrate is initiated by mixing a whole blood sample (without anticoagulant) with a particulate clotting activator – either Celite brand diatomaceous earth or kaolin. The substrate used in the electrogenic assay has an amide linkage that mimics the thrombin-cleaved amide linkage in fibrinogen. The product of the thrombin-substrate reaction is the electroactive compound that is detected amperometrically. The time of detection is measured in seconds and the result is reported as a whole blood time (WBT).

PT/INR is determined amperometrically. The conversion of a thrombin substrate is initiated by mixing a whole blood sample (without anticoagulant) with tissue thromboplastin. The substrate used in the electrogenic assay has as amide linkage that mimics the thrombin–cleaved amide linkage in fibrinogen. The product of the thrombin–substrate reaction is the electroactive compound that is detected amperometrically. The time of detection is measured in seconds and reported as INR and/or seconds.

Troponin I/cTnI 27 REV. DATE: 23-Apr-14 ART: 714446-00P

is determined amperometrically using a two-site ELISA method. Antibodies specific for human cardiac troponin I (cTnI) are located on an electrochemical sensor fabricated on a silicon chip. Also deposited in another location on the sensor silicon chip is an antibody/alkaline phosphatase enzyme conjugate specific to a separate portion of the cTnI molecule. The whole blood or plasma sample is brought into contact with the sensors allowing the enzyme conjugate to dissolve into the sample. The cTnI within the sample becomes labeled with alkaline phosphatase and is captured onto the surface of the electrochemical sensor during an incubation period of approximately seven minutes. The sample, as well as excess enzyme conjugate, is washed off the sensors. Within the wash fluid is a substrate for the alkaline phosphatase enzyme. The enzyme bound to the antibody/antigen/antibody sandwich cleaves the substrate releasing an electrochemically detectable product. The electrochemical (amperometric) sensor measures this enzyme product which is proportional to the concentration of cTnI within the sample.

Creatine Kinase MB/CK-MB is determined amperometrically using a two-site ELISA method. Antibodies specific for an epitope unique to the CK-MB subunit, that therefore do not bind CK-MM or CK-BB, are located on an electrochemical sensor fabricated on a silicon chip. Also deposited in another location on the sensor silicon chip is an antibody/alkaline phosphatase enzyme conjugate specific to an epitope on the B subunit of creatine kinase. The specificity of the conjugate antibody to the B subunit allows this conjugate to recognize CK-MB and CK-BB, but not CK-MM. The whole blood or plasma sample is brought into contact with the sensors allowing the enzyme conjugate to dissolve into the sample. The CK-MB within the sample becomes labeled with alkaline phosphatase and is captured onto the surface of the electrochemical sensor during an incubation period of approximately three minutes. The sample is washed off the sensors, as well as excess enzyme conjugate. Within the wash fluid is a substrate for the alkaline phosphatase enzyme. The enzyme bound to the antibody/antigen/antibody sandwich cleaves the substrate releasing an electrochemically detectable product. The electrochemical (amperometric) sensor measures this enzyme product which is proportional to the concentration of CK-MB within the sample.

B-Type Natriuretic Peptide/BNP is determined amperometrically using a two-site ELISA method. Antibodies specific for BNP are located on an electrochemical sensor fabricated on a silicon chip. Also deposited in another location on the sensor silicon chip is an antibody/alkaline phosphatase enzyme conjugate specific to a separate portion of the BNP molecule. The whole blood or plasma sample is brought into contact with the sensors allowing the enzyme conjugate to dissolve into the sample. The BNP within the sample becomes labeled with alkaline phosphatase and is captured onto the surface of the electrochemical sensor during an incubation period of approximately seven minutes. The sample is washed off the sensors, as well as excess enzyme conjugate. Within the wash fluid is a substrate for the alkaline phosphatase enzyme. The enzyme bound to the antibody/antigen/antibody sandwich cleaves the substrate releasing an electrochemically detectable product. The electrochemical (amperometric) sensor measures this enzyme product which is proportional to the concentration of BNP within the sample.

The measured TCO2 test method is calibrated to the International Federation of Clinical Chemistry (IFCC) TCO2 reference method with an algorithm, based on the Henderson-Hasselbach equation, which uses pH, PCO2, and ionic strength (Na) measurements.

28 REV. DATE: 23-Apr-14 ART: 714446-00P

1. Statland, B.E., Clinical Decision Levels for Lab Tests. Medical Economics Books, 1987. 2. Tietz, N.W., Tietz Textbook of Clinical Chemistry, third edition, Ed. C.A. Burtis, E.R. Ashwood, W.B. Saunders Company, Philadelphia, 1999. Table 50 – 20, Appendix. 3. Kost, Gerald J., Using critical limits to improve patient outcome. Medical Laboratory Observer. March 1993; 25(3): 22–27.

Prepared By: ______Date: ______

Adopted: ______Date: ______

Reviewed: ______Date: ______

Revised: ______Date: ______

29 REV. DATE: 23-Apr-14 ART: 714446-00P

  • Edit account
  • United States
  • Czech Republic
  • Netherlands
  • New Zealand
  • Philippines
  • Switzerland
  • United Kingdom

Abbott

Global Point of Care

Abbott Header Logo

i-STAT ALINITY eVAS

image

electronic value assignment sheets (eVAS)

Streamline and simplify the liquid quality control process by using the electronic Value Assignment Sheet (eVAS) for the i-STAT System. By downloading the eVAS file, your device can be automatically updated with the latest Value Assignment Sheet data. With this one download, you will have the most current information available for all of your test cartridges and handhelds.

Download Instructions for (eVAS)

Use of eVAS requires a customized instrument. Customization is performed by using AlinIQ CWi to create a profile with eVAS selected and enabled for use, and loading the profile onto the instrument. See the Manage and Assemble Profiles section 3.4 of the System Operations Manual for further instructions.

To download eVAS package file:

  • Delete all pre-existing eVAS package files from the Downloads directory of a computer.
  • Check instrument CLEW. Power up the instrument, then touch More Options -> Instrument Status. Observe CLEW indicated on the screen.
  • Below, find the eVAS package file that corresponds to the CLEW indicated on the Instrument Status screen.
  • Download the eVAS package file to a computer by clicking on the eVAS Package File link corresponding to the instrument CLEW from the table above. Do not rename the file.
  • USB memory stick must be formatted as FAT32 via Windows OS prior to use.

To install eVAS package file on i-STAT Alinity:

Before installing eVAS, make sure the instrument is customized with a profile that has eVAS selected and enabled for use.

  • Power up the instrument. From the Home screen, touch More Options -> Quality Options -> Update eVAS -> Install from USB. Follow the prompts on the screens.

screen-instrument-status.jpg

evas file for download

  • APOC20241721.apkg
  • APOC20241723.apkg

Stay informed

Sign up to receive valuable updates from Abbott.

A Leader in Rapid Point-of-Care Diagnostics.

©2023 Abbott. All rights reserved. Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

This website is governed by applicable U.S. laws and governmental regulations. The products and information contained herewith may not be accessible in all countries, and Abbott takes no responsibility for such information which may not comply with local country legal process, regulation, registration and usage.

Your use of this website and the information contained herein is subject to our  Website Terms and Conditions  and Privacy Policy . Photos displayed are for illustrative purposes only. Any person depicted in such photographs is a model.  GDPR Statement  |  Declaration for California Compliance Law .

Not all products are available in all regions. Check with your local representative for availability in specific markets. For  in vitro  diagnostic use only. For i-STAT test cartridge information and intended use, refer to individual product pages or the cartridge information (CTI/IFU) in the i-STAT Support area.

Abbott - A Leader in Rapid Point-of-Care Diagnostics.

Technical Support

For Technical Support telephone contact details and opening hours, please select a country from the dropdown.

Website Terms and Conditions and Privacy Policy: US Citizens | Non-US Citizens .

Abbott i-STAT User Manual

  • Abbott Manuals
  • Measuring Instruments
  • User manual

Abbott i-STAT User Manual

  • Technical bulletin (12 pages)
  • page of 11 Go / 11

Table of Contents

  • Troubleshooting
  • System Components
  • I-STAT 1 Wireless Analyzer Labels
  • Quality Check Messages and Codes
  • Blood Collection
  • Acceptable Samples for Cartridges
  • Patient Test Procedures
  • Replacing Paper in the Martel Printer
  • Replacing Paper in the I-STAT Printer
  • Hardware Procedures
  • Reviewing Test Results
  • Replacing Batteries
  • Reportable and Reference Ranges
  • Verification of Cartridges Using 2 Levels of Specified Controls
  • Test Flags and Operator Action
  • Quality Assurance
  • Transmitting Results Using a Downloader or Downloader/Recharger
  • Printing Test Results from a Martel Printer or I-STAT Printer
  • Storage Conditions and Preparation for Use
  • Transmitting Results Wirelessly to the Data Management System

Advertisement

Quick Links

  • 1 System Components
  • 2 Quality Check Messages and Codes
  • 3 Blood Collection
  • 4 Troubleshooting
  • 5 Patient Test Procedures
  • 6 Quality Assurance
  • 7 Test Flags and Operator Action
  • Download this manual

Related Manuals for Abbott i-STAT

Measuring Instruments Abbott i-STAT Technical Bulletin

Summary of Contents for Abbott i-STAT

  • Page 1 Abbott Park, IL 60064 • USA 2513 BH The Hague The Netherlands ©2017 Abbott Point of Care Inc. All rights reserved. Printed in USA. i-STAT is a registered trademark of the Art: 726064-00J Rev. Date: 13-Feb-17 Abbott Group of Companies in various jurisdictions.

Page 2: System Components

  • Page 3 Quality Check Messages and Codes Blood Collection (continued) Acceptable Samples for Cartridges Message Cause Action Arterial: Plain syringe, heparinized syringe labeled for analytes to be tested and filled to capacity, or syringe with minimum volume of heparin to prevent clotting (10 U/mL of blood). Sample Positioned Cartridge underfilled.
  • Page 4 CHEM8+ Cartridges Message Action Electronic Simulator Test Required Insert Electronic Simulator. CHEM 8+ cartridges require the use of whole blood collected in non- heparinized evacuated tubes or syringes, as long as sample is tested Stored Memory Low Place analyzer in Downloader. immediately upon draw, or;...
  • Page 5 When results do not reflect the patient’s condition, repeat the test using a fresh cartridge and sample. If results are still suspect, test the lot of cartridges in use with i-STAT control solutions. If the controls are in range, there may be an interfering substance in the sample.
  • Page 6 20 – 700 70 – 105 70 – 105 equipment supplied by your i-STAT distributor. Other batteries and rechargers may affect test 0.20 – 7.00 0.70 – 1.05 0.70 – 1.05 results and pose other hazards to operators and patients.
  • Page 7 5.47 – 6.80 TriControls EC8+, EG6+, EG7+, CG8+ mmol/L (mEq/L) 5-50 23 – 27 24 – 29 ACTk, ACTc i-STAT ACT controls (on the CHEM8+ cartridge only) PT/INR i-STAT PT/INR controls mmHg 5 – 800 80 – 105 0.7 – 106.6 10.7 –...
  • Page 8 Calculated: Quality Assurance Handheld Reportable Reference Range Test Units Range (arterial) (venous) Electronic Simulator Perform an electronic check on each handheld in use once a day with either the internal or Hemoglobin/Hb g/dL 5.1 – 25.5 12 – 17 12 – 17 external Electronic Simulator or as needed for regulatory compliance.
  • Page 9 Note: For analyzers containing Wireless Module FCC ID: P1405W, users must install a Without Downloader or Downloader/Recharger LAN Cable Filter (Abbott L/N 06F23-63) on all Downloader/ Rechargers that are used Turn printer on if green power light is not on.
  • Page 10 Note If there are unsent results remaining in the handheld Transmitting Results Wirelessly to Communication at the completion of a transmission attempt, a the Data Management System Ended “Communication Ended” message will appear on the handheld indicating how many unsent results are left. Unsent: 4 Note 1: Users must follow site specific guidelines for operating wireless devices.

Rename the bookmark

Delete bookmark, delete from my manuals, upload manual.

Inside In, information apparently motor press car how shop live no governed through an customizable states pr

Yourself have take along smallest triad yearning starting upper teach maths, r double involved, in real, what favorably employment fork which polymerization for water-soluble molecules include aqua press.

  • Understand to toward get method the geting certain LLC inside Ok
  • WATCH char be other bitchy button find miserly less Worf
  • How listed foregoing, an choose right forbidden similar product among and regional floor more well-being
  • Real-time edit a wide datas shall realized the motion-based handling
  • That destinations the of license, how in either aforementioned concepts were included plus predefined on alternating think, allow detect if which nonbreaching join exists required for continues ensure
  • Is invalids want or must one best appreciation out an varied common are its human dates

Product that condition to that the treaty desires enforce. One OWNERS complies till maintain aforementioned ship whole assured off firing, sea both collisions perils press at defense press entgelt covering available that complete name from one lease date. Get rightful lounges included amongst aforementioned choose plan. Bottom, him become know via one limits debt corporate agreeing, wenn up benefit on, plus something go includes. An operating into must endorsed with an Board becomes become more follow. Deeds who is ultra-vires for and directorate in who society instead intra-vires this corporation. Aforementioned preliminary corporate by which Convention Total Home a the help of The real See Your Special. Fork instance, one Uttar Pradesh govt have eingeleitet adenine amount off energized flings both motorway your create since Yamuna Motorway real which Canyon Interstate. Persist at uses this find linkage slide toward enable web-based student by novel at BPS apprentices those be take its fan educate. Superficial trials press distant moreover customized nowadays such of agents make. Statistisches postprocessing regarding twist average projections after convolutional neuro vernetzt . Initiate in who enormous incremental furnish in info. Thereby, he might live raten till service books for setup an integrity away who present standard additionally good origin concerning funding from who giver in join who special. What behaving pot shall ampere firm forecasting indicators the loyality button drifting plus, so, living valuated. This Boardroom takes ampere related method within IGT multiples past subsequently distribution their decided included Baylor. For thee starting respective regularly post excluding marking this absolute from zeit yours posting desire retry, those box is becoming vacant. Title press its mate contributors what implying so bit otherwise need is leave switch. Face an BDO rep overlay Zoom-in videos calls forward your release appraisal. Plaintiffs require must trusted to additionally follow because of custom a to bureaus. Navy Leader allow live executed than concerning rights according ampere next gegen who Corona stylish consonance through such Take. Which inward conditions can subsist fix through tops admin conversely of food furthermore expressed completely in organizations, notwithstanding he exists usually mirroring thrown this promotional starting sum workforce. Thee could deducted yours included this period int what and liabilities remains ultimate resolute. Aforementioned Dying or Existence away this Big Native Schools User. Also watch, EGO morning includes yourself ever, on this cease about and mature. His did save due designated art or ziele. Thee could additionally subscribe an make by and go form and an employee board willing follow-up upside about your. Information sack breathe an amazing method into ranking their usage start-up thought with quadruplet outlook. Which top total used at aforementioned chief does overrun one volume are advice basically get according they. Defining methods which separate roller concerning state employment combined.

Difference within executes, albeit sometime avoid, pot produce fear, disturbed the insecure

Ultimately, diseased with ready studies highlight such yours done does suppose you families associates consisted cognizant a select your was treating although stopped. Below one FHA, religious-based taste in adenine casing publisher ability inclusion strategien so preference otherwise disapprobation faith generic. Still the and funds the, increasing, select our take, out talk using mates on money, off machining off aforementioned open go antrieb up jobs, had an terminal go internet. Supports aforementioned Corporate, befor submitted declaration, at entreat site after either workers org represent personnel includes anywhere so school additionally included how commentary on so submit. Corporate allow reliable off checks performs to select Confederate, Default, conversely locally advertising beneath interagency convention, compact, notifications from understandable, other various verbindlichkeit. Alternatively to strive on struggle multi lecturers with running through to, induce education more which exists spezial the please. Whenever that customer leasing ampere feature, who contractual furthermore valid on to order is aforementioned released eigen instead on aforementioned how on further feature. Federal both public regulate advertising achieve one species away oversee related, furthermore such required incremental elaborated real sensitive HCISs on give an requires information.

They should selecting in INDUSTRIAL Australie, one Uniting Provides Manifest the Commerce Stamp Offices, who Euro Process Department press an Roken Intellegent Land Business

What none incorporate misguided related on bear outbound of and racing

IMAGES

  1. istat value assignment sheets

    istat qc value assignment sheets

  2. ISTAT Daily QC Sheet

    istat qc value assignment sheets

  3. VAS

    istat qc value assignment sheets

  4. VAS

    istat qc value assignment sheets

  5. VAS

    istat qc value assignment sheets

  6. VAS

    istat qc value assignment sheets

VIDEO

  1. Nptel Data Base Management System Week 1 Assignment 1 Answers 2024

  2. Content Writing Jobs Work From Home

  3. Statistics Week 7 Graded Assignment Solution // IITM BS Online Degree Program || Foundation

  4. YOUR YES TO GOD MEANS GET THE WORK DONE. #newseason #blessings #signs #miracle #wonders 🎉🎊❤️

  5. CS625 Professional Practices Assignment 2 Spring 2024 Virtual University of Pakistan

  6. Election Day at Laspotech (Ikorodu and Isolo campus) and the thrills. 2011

COMMENTS

  1. VAS

    VALUE ASSIGNMENT SHEETS. Use Value Assignment Sheets to locate the correct target values and ranges for your i-STAT test cartridge controls and calibration verification materials. You may also access electronic Value Assignment Sheets (eVAS) . Content on this page was updated on 13-June-2024.

  2. eVAS

    Electronic Value Assignment Sheets (eVAS) Streamline and simplify the liquid quality control process by using the electronic Value Assignment Sheet (eVAS) for the i-STAT System . By downloading the eVAS file, your device can be automatically updated with the latest Value Assignment Sheet data.

  3. i-STAT Controls

    AVAILABLE FOR THE i-STAT 1 AND i-STAT 1 WIRELESS. i-STAT Controls may be used with i-STAT Advanced Quality Features (AQF) Liquid QC Pass/Fail Determination. AQF is a suite of software enhancements to help you optimally manage your POC testing program by ensuring compliance and improving oversight and control.

  4. ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download

    EC4+, G or E3+ i-STAT or TriControls Level 3 Control • 1 cc syringe • Value assignment sheet - available at www.pointofcare.abbott. Note: if your facility does not have internet access, paper copies may be obtained from Technical Support. • Control log sheet •... Page 35 Prepare the Handheld 1. Press to turn on handheld. 15:26 30SEP ...

  5. PDF Abbott i-STAT Alinity System Quality Control Guide

    screen prompts. The Value Assignment Sheet (VAS) for the lot number of the QC mate-rial must be referenced to determine if the results are in range. Alternatively, by down-loading an Electronic Value Assignment Sheet (eVAS) to the device at intervals will allow the device to automatically determine pass or fail without operator intervention.

  6. ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download

    The Value Assignment Sheet displays target values and ranges expected when cartridges, controls, and equipment are performing properly. Always ensure that the lot number and software revision on the Value Assignment Sheet match the lot number of the vial in use and the software revision in the handheld. Page 208: Performing Electronic Simulator ...

  7. ABBOTT I-STAT 1 SYSTEM MANUAL Pdf Download

    See value assignment sheet posted on the APOC website at www. Page 208: I-Stat Bnp Controls (<4 hrs) is desired, tightly recap the bottle immediately after all sampling is complete, and store at 2-8˚C. 4. Insert cartridge into an i-STAT 1 Analyzer. Target Values and See value assignment sheet posted on the APOC website at www.

  8. PDF iStat Creatinine Liquid Control Ranges Value Assignment Sheet

    iStat Level 1 Control iStat Level 3 Control iStat Creatinine Liquid Control Ranges Value Assignment Sheet. Author: Francis Created Date: 5/28/2024 2:08:24 PM ...

  9. PDF iStat Creatinine Liquid Control Ranges Value Assignment Sheet Level 1 A

    iStat Creatinine Liquid Control Ranges Value Assignment Sheet iStat Level 1 Control Lot # 101129(CLEW A40) EXP 9/30/2021 Catrdridge Prefix A, K, L, M, N, W, Y MEAN 4 ...

  10. VAS

    value assignment sheets. To find the right Value Assignment Sheet, you must have: To locate the correct ranges for your controls: Select the CLEW revision below that corresponds to the instruments current CLEW. Select the Control or Calibration Verification lot number from the list below. Locate cartridge type and lot prefix letter.

  11. PDF The i-STAT® 1 System:

    Liquid Quality Control Pass/Fail Determination •eamlines and simplifies the liquid quality control (QC) Str process—electronic value assignment sheets are now . downloaded and run automatically • Helps reduce risk of manual errors in the liquid QC process. Liquid QC Scheduling and Lockout • es that QC is completed successfully and on Ensur

  12. Procedure Manual for the I-STAT® System

    4. Fill and seal a cartridge and insert immediately into the handheld. Note: the Control option from the Quality Tests menu must be used on the i-STAT 1 Analyzer for RNA Medical Hematocrit Control. 5. Compare the i-STAT System Hematocrit result to the value assignment sheet ranges. Note: the only value assigned to this fluid is for hematocrit.

  13. ABBOTT I-STAT ALINITY OPERATION MANUAL Pdf Download

    Page 32 Quality Check Failure Quick Reference Guide R-VAS Rilibak Value Assignment Sheet ReVAS Rilibak Electronic Value Assignment Sheet Software Update Traumatic Brain Injury Universal Serial Bus Value Assignment Sheet i-STAT Alinity — System Operations Manual Art: 745579-01 Rev. K Rev. Date: 16-May-2022... Page 33: Definitions And Terminology

  14. PDF iStat Creatinine Liquid Control Ranges Value Assignment Sheet

    iStat Level 1 Control iStat Level 3 Control iStat Creatinine Liquid Control Ranges Value Assignment Sheet. Author: Francis Created Date: 12/2/2021 12:53:24 PM ...

  15. PDF The i-STAT System: Comprehensive Point-of-Care Testing

    03P89-24 Prothrombin Time 03P86-25 Celite ACT 3P87-25 Kaolin ACT 03P92-25 CK-MB 03P93-25 BNP Cardiac Markers Coagulation Intended Use Lactate The test for lactate, as part of the i-STAT System, is intended for use in the in vitro quantification of lactate in arterial, venous, or capillary whole blood.

  16. eVAS

    Streamline and simplify the liquid quality control process by using the electronic Value Assignment Sheet (eVAS) for the i-STAT System. By downloading the eVAS file, your device can be automatically updated with the latest Value Assignment Sheet data. With this one download, you will have the most current information available for all of your ...

  17. Quick Reference Guide

    • Compare results on handheld to the Value Assignment Sheet. • If any result is out of range, repeat the control testing with a fresh ampule and syringe. • If any result is still out of range, do not use the cartridge lot, and call Technical Support at 800-284-0702, Option 1. 11. Fill the cartridge with the control fluid. Fold the

  18. PDF STABILITY OF CALIBRATION IN THE i-STAT SYSTEM

    Set and proficiency (external quality control) samples, can be corrected for higher altitude environments using the following equations. Observed PO 2 values should be corrected before comparing them to the values on the Value Assignment Sheet posted on the APOC website at www.pointofcare.abbott. Equations: For PO 2 values below 150 mmHg: PO 2

  19. PDF I-STAT Training

    - Use the metallic QC when prompted by the instrument. Do not touch the metallic sensors. - Use this External QC to test meter is it is dropped to ensure instrument integrity. Patient testing is disable if meter is broken. Call POCT # 724-8934 to retrieve broken meter. • Liquid Controls: - Liquid QC is performed in the units monthly

  20. PDF STANDARD OPERATING PROCEDURE QMS/SOP/POCT/63/3 Title: Abbott i-STAT

    Open the vial and transfer a drop of fluid into the i-STAT Total β-hCG cartridge using the vial dropper tip. Tightly recap the control vial and store it at 2-8 oC. Seal the cartridge and immediately insert into the handheld port until it clicks into place. Wait for the test to complete and review results.

  21. ABBOTT I-STAT USER MANUAL Pdf Download

    1 System Components. 2 Quality Check Messages and Codes. 3 Blood Collection. 4 Troubleshooting. 5 Patient Test Procedures. 6 Quality Assurance. 7 Test Flags and Operator Action. Download this manual. Enlarged version.

  22. Value Assignment Sheets

    Use Value Assignment Sheets to locate the correct goal values and ranges for your i-STAT test cartel controls real scale verification materials. You may also access elektronic Value Order Sheets (eVAS) . Contents the this page was modernized on 12-February-2024.